Accuracy and versatility makes the HF LCR Meters the ideal choice
for many different tasks and applications in manufacturing and test.
The wide range of frequency specifications means that a model can be
selected that best meets application requirements and within budget.
Measurement parameters
Capacitance
(C) Inductance
(L) Resistance
(R) Reactance
(X) Conductance
(G) Susceptance
(B) Dissipation Factor
(D) Quality Factor
(Q) Impedance
(Z) Admittance
(Y) Phase Angle (Ø)
High measurement accuracy
Capacitance, inductance and impedance basic accuracy is an
excellent ±0.05%. Dissipation factor accuracy is ±0.0005 and the
quality factor accuracy is ±0.05%.
Variable drive and bias levels
AC drive levels up to 1V or 20mA can be selected to evaluate
components in realistic operating environments.
The /D1 DC bias option provides 0 to +40V dc bias voltage and 0
to +100mA dc bias current while the /D2 option provides -40V to +40V
dc bias voltage.
External control
The GPIB interface is used to control the instrument and read
back measured values for applications such as quality control or for
archiving purposes.
An Ethernet interface similarly allows the instrument to be
controlled and to send out data - allowing it to be integrated into
many test environments.
Wide range of interfaces
An external monitor or projector may be connected to the
instrument's VGA output. The ability to provide a large screen
display is invaluable in production environments or for teaching and
training.
Instrument control from both a keyboard and mouse is available.
Any keyboard or mouse, with either PS/2 or USB interfaces, can be
simply connected to provide an alternative method of instrument
control and operation.
Data storage and retrieval
All measurement and setup data can be stored using the Ethernet
interface or a USB flash memory (supplied as standard).
Setup dataUp to 20 instrument setups may be stored locally.
Bin handling
The /B1 option (non-isolated 5V) or /B2 option (isolated 24V)
signals are available through a 25-way D-type connector. 10 bins can
be set using absolute or percentage limits.
Printer outputs
Hard copy printouts can be obtained in a number of ways including
direct to an HP-PCL compatible graphics printer or Epson compatible
text/ticket printer. A networked HP-PCL compatible printer may
also be used via Ethernet connection.
Component connections
Four front panel BNC connectors permit three or four terminal
connections with the screens at ground potential.
The 1J1011 component fixture, supplied as standard, ensures
optimum performance when measuring a wide range of components of
leaded components and devices.
1J1012 (2 Terminal) and 1J1014 (4 Terminal) fixtures allow
connection to surface mount devices.
Protection against charged capacitors
High precision measuring instruments can be damaged by charged
capacitors which can cause costly repairs and unacceptable downtime.
All the models in the range incorporate protection against charged
capacitors.
Comprehensive and precise component tests at higher
frequencies
The 6500P series capabilities are best summarised by
“Comprehensive and precise component tests at higher frequencies”.
The instruments are the perfect solution for those who have
demanding component measurement needs.
Technical Specifications
Specification applies to all LCR meters in the range unless a
model number or option is specified.
Measurement parametersAny of the following
parameters can be measured and displayed:
AC Functions Capacitance
(C) Inductance
(L) Resistance
(R) Reactance
(X) Conductance
(G) Susceptance
(B) Dissipation Factor
(D) Quality Factor
(Q) Impedance
(Z) Admittance
(Y) Phase Angle (Ø)
Display formatSeries or parallel equivalent circuit - all
parameters
Test conditions
Frequency range6505P 20Hz to
5MHz 6510P 20Hz to 10MHz 6515P 20Hz
to 15MHz 6520P 20Hz to 20MHz 6530P
20Hz to 30MHz 6550P 20Hz to 50MHz 65120P 20Hz to
120MHz
Frequency step size: < 1mHz Accuracy of set frequency
±0.005%
AC Drive level10 mV to 1V rms* 200 µA to 20 mA
rms* *Varies with frequency
Signal source impedance: 50 Ω
nominal
DC Bias (option) /D1 0 to +100mA dc bias current and 0 to +40V
dc bias voltage /D2 -40V to +40V dc bias voltage
Binning (option) 10 bins with absolute and percentage
limits 25 way D-type interface connector
/B1
(Non-Isolated) Common 0V. Bin outputs 0 to 5V (nominal) with
>10mA current sink capability /B2 (Isolated) Common 24V
input. Outputs 0 to 24V with >10mA current source capability.
Setup dataUp to 20 instrument setups may be stored locally.
Measurement connectionsFour front panel BNC
connectors permit two, three or four terminal connections with the
screens at ground potential.
1J1011 Component Fixture
(supplied as standard) allows connection to leaded components and
devices.
1J1012 (2 Terminal) and 1J1014 (4 Terminal) fixtures allow
connection to surface mount devices.
Measurement accuracy Dissipation factor ±0.0005
(1+D2)**
Quality factor ±0.05 %(
Q+1/Q)**
Capacitance / Inductance /
Impedance ±0.05%**
*Varies with frequency, drive level and
measured impedance
General
Power supplyInput voltage 90 V AC to 264 V AC (Autoranging)
Mains frequency47 to 63 Hz
Display8.4″ VGA (640 x 480) colour TFT with touch panel
Local printerHP-PCL compatible graphics
printing Centronics / parallel printer port, Epson compatible
text / ticket printing
Network printerHP-PCL compatible graphics printing
Remote triggerRear panel BNC with internal pull-up,
operates on logic low or contact closure
GPIB interfaceExternal instrument control. 24 pin IEEE 488
connector
USB interfaceTwo Universal Serial Bus Interfaces USB 1.1
compliant
VGA interface15-way D-type connector to drive an external
monitor in addition to the internal screen
Network interface10/100-BASETX Ethernet controller. RJ45
connector
Keyboard interfaceStandard USB or PS/2 keyboard port.
Instrument front panel remains active with keyboard plugged in.
Mouse interfaceStandard USB or PS/2 mouse port. Touch
screen remains enabled when the mouse is connected.
Bin handler (option)/B1 Isolated (24 V) and /B2
non-isolated (5 V). 25-way D-type connector
Environmental conditions
This equipment is intended for indoor use only in a non-explosive
and non-corrosive atmosphere
Temperature RangeStorage -20°C to 60°C Operating 0°C to
40°C Full Accuracy 18°C to 28°C
Relative humidityUp to 80% non-condensing
AltitudeUp to 2000m
Installation categoryII in accordance with IEC664
Pollution degree2 - mainly non-conductive
SafetyComplies with the requirements of EN61010-1
EMCComplies with EN61326 for emissions and immunity
MechanicalHeight 190 mm (7.5″) Width 440 mm
(17.37″) Depth 525 mm (20.5″) Weight 13.8 kg (32 lb)
Order codes and options
6505P 5 MHz Precision Impedance
Analyzer |
|
1J6505P |
|
6510P 10 MHz Precision Impedance
Analyzer |
|
1J6510P |
|
6515P 15 MHz Precision Impedance
Analyzer |
|
1J6515P |
|
6520P 20 MHz Precision Impedance
Analyzer |
|
1J6520P |
|
6530P 30 MHz Precision Impedance
Analyzer |
|
1J6530P |
|
6550P 50 MHz Precision Impedance
Analyzer |
|
1J6550P |
|
65120P 120 MHz Precision Impedance
Analyzer |
|
1J65120P |
|
|
All units are supplied with user
manual, 2 m AC power cable, universal component fixture
(1J1011) and USB memory |
|
Options |
Bin handler |
|
/B1 |
Bin handler ( isolated 24V ) |
|
/B2 |
DC Bias ( 0 to +100mA and 0 to +40V ) |
|
/D1 |
DC Bias ( -40V to +40V ) |
|
/D2 |
|
Optional Accessories |
Surface Mount Fixture ( 2 Terminal ) |
|
1J1012 |
Surface Mount Fixture ( 4 Terminal ) |
|
1J1014 |
Certificate of Calibration with results |
|
1JCALRES |
|
|
Wayne Kerr’s policy is one
of continuous development and consequently the product may
vary in detail from the description and specification in this
publication. | |